CVE-2025-14095
中文标题:
(暂无数据)
英文标题:
Privilege boundary violation in Radiometer Products
漏洞描述
中文描述:
(暂无数据)
英文描述:
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
CWE类型:
标签:
受影响产品
| 厂商 | 产品 | 版本 | 版本范围 | 平台 | CPE |
|---|---|---|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | All application software versions with Windows 7, Windows XP as underlying OS | - | - |
cpe:2.3:a:radiometer_medical_aps:abl90_flex_and_abl90_flex_plus_analyzers:all_application_software_versions_with_windows_7,_windows_xp_as_underlying_os:*:*:*:*:*:*:*
|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Application software versions < 3.5MR11 with Windows 10 as underlying OS | - | - |
cpe:2.3:a:radiometer_medical_aps:abl90_flex_and_abl90_flex_plus_analyzers:application_software_versions_<_3.5mr11_with_windows_10_as_underlying_os:*:*:*:*:*:*:*
|
| Radiometer Medical Aps | AQT90 FLEX Analyzers | All Application software versions <= 8.13 MR2 | - | - |
cpe:2.3:a:radiometer_medical_aps:aqt90_flex_analyzers:all_application_software_versions_<=_8.13_mr2:*:*:*:*:*:*:*
|
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS | - | - |
cpe:2.3:a:radiometer_medical_aps:abl800_basic_and_abl800_flex_analyzers:application_software_versions_<_6.20_mr2_with_windows_7,_windows_xp_as_underlying_os:*:*:*:*:*:*:*
|
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | Application software versions < 6.20 MR2 with Windows 10 as underlying OS | - | - |
cpe:2.3:a:radiometer_medical_aps:abl800_basic_and_abl800_flex_analyzers:application_software_versions_<_6.20_mr2_with_windows_10_as_underlying_os:*:*:*:*:*:*:*
|
| Radiometer Medical Aps | ABL9 Analyzers | Application software versions < 1.5.0 | - | - |
cpe:2.3:a:radiometer_medical_aps:abl9_analyzers:application_software_versions_<_1.5.0:*:*:*:*:*:*:*
|
解决方案
中文解决方案:
英文解决方案:
临时解决方案:
参考链接
cve.org
CVSS评分详情
3.1 (cna)
MEDIUMCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
时间信息
利用信息
数据源详情
| 数据源 | 记录ID | 版本 | 提取时间 |
|---|---|---|---|
| CVE | cve_CVE-2025-14095 |
2025-12-19 03:24:27 | 2026-01-12 02:11:06 |
| NVD | nvd_CVE-2025-14095 |
2025-12-19 03:25:39 | 2026-01-12 02:27:30 |